Sanyo MCO-18AIC Service Manual page 45

Table of Contents

Advertisement

<
When stacking of combination as shown in Fig.2.
MCO-17AIC
MCO-17AC
MCO-15AC
MCO-18AIC
1. Turn off the power switch and disconnect a plug of each unit.
2. Check that the lower unit is level.
3. Apply the protective sticker enclosed in this kit at each corner on the top of the lower unit to avoid
scratches or damage.
4. Stack the unit so that both units can be aligned straight. Also check the upper unit is level. If it is not
level, keep the unit even by adjusting the leveling legs.
5. Remove 2 hooks on the rear side of the lower unit by unscrewing each 2 fixing screws.
6. Remove 1 screw on the bottom right and left on the rear side of the upper unit.(only MCO-18AIC)
7. Fix the stacking plate B at the right and left on the rear of the lower and upper unit with 3 screws
removed in step 5 and enclosed in this kit.
8. Fix the stacked unit to the wall with 2 hooks on the rear of the upper unit and rope or chain.
Protective
sticker
WARNING
Select a level and sturdy floor having enough strength for installation of stacked module.
Never stack 3 or more units.
Take care not to drop or tip over the unit when stacking as this can cause injury or damage
of the unit.
MCO-18AIC
MCO-17AIC
MCO-17AC
MCO-15AC
Fig.2
Stacking plate B
Fig.3
42
>
MCO-17AIC
MCO-17AC
MCO-15AC
MCO-17AIC
MCO-17AC
MCO-15AC
Stacking plate A is not used.
Hook
7FB6P104038000

Hide quick links:

Advertisement

Table of Contents
loading