Sanyo MCO-18AIC Service Manual page 21

Table of Contents

Advertisement

Followings are initial values in non-volatile memory:
Internal temperature SV
CO
level SV
2
Balance value of humidifying heater (F06)
Balance value of door heater (F07)
CO
zero adjustment data (F10)
2
CO
span data (F11)
2
Internal temperature zero adjustment data (F15)
Key Lock data
Auto Changer data (the 3
Demo Mode data (the 2
Auto Zero (the 1
Input "F13" on temperature display and press ENT key to display "000" on
CO
display, then change the 1
2
CO
zero adjustment data and CO
2
Note)
F14: Check the performance for LED, drivers
<Direction for
Input "F14" on temperature display and press ENT key to perform with lamp
use>
and the all digits flashing at every 0.5seconds intervals. Press ENT key to
stop performance.
If there is no key operation for approx. 90seconds, automatically comes to
end
F15: Display an adjustment value for temperature zero point (initial: 0.8 C)
<Direction for
Input "F15" on temperature display and press ENT key to display an
use>
adjustment value on CO
key or leave the device for 90seconds, automatically returns to PV display.
(Ex.) In "37.0" is displayed on the control panel and actual temperature is "37.8", if
you perform temperature calibration (actual temperature "37.8" is input),
"+0.8" would be housed in F15. In Function Mode, input F15 and press ENT
key to display the value on CO
F21: Set serial communication ID (initial: 000)
<Direction for
Input "F21" on temperature display and press ENT key to show "000" on CO
use>
display. Then input the value that identifies the product (serial communication
ID Number) and press ENT key to set serial communication ID. The range is
000~255. (000: communication OFF)
F22: Set serial communication mode (initial: 000)
<Direction for
Input "F22" on temperature display and press ENT key to show "000" on CO
use>
display. Then change the 3
Control mode (the 3
Baud rate (the 2
F24: Interlock the remote alarm with buzzer (initial: 000)
<Direction for
Input "F24" on temperature display and press ENT key to show "000/001" on
use>
CO
display. Then change the 1
2
000: Buzzer/Remote alarm not interlocked
001: Buzzer/Remote alarm interlocked
rd
digit in F08)
nd
digit in F08)
st
digit in F08)
Service code should not be initialized.
display for temperature zero point. If you press ENT
2
rd
digit to set the communication mode.
rd
digit)
nd
digit)
18
: 37.0 C
: 0.0%
: 12
: 4
: 5000
: 5000
: 008
: Key Lock OFF
: Auto Changer OFF
: Demo Mode OFF
: Auto Zero ON
st
digit to "2" and press ENT key to initialize
span data in non-volatile memory.
2
display.
2
0: Local (initial)
1: Remote
0: 2400bps
1: 4800bps
2: 9600bps
st
digit to "000" to set the interlock mode.
2
2