MultiFiber Pro Getting Started Guide
2
While LOSS LIMIT shows, press
or increase the limit value. To change the value quickly, hold
down the key.
3
To save the setting and exit setup mode, press
hold down
Measure Loss
1
Set the reference as described on page 11.
2
Make sure the limit is correct. To see the limit, hold down
for 2.5 seconds, press
LOSS LIMIT, then press
3
Clean and inspect the connectors on the link and on the
required test cords.
4
Disconnect the test cord from the meter, then make the
appropriate connections. The figure on page 15 shows
connections for a link with unpinned connectors. The Users
Manual shows connections for other types of links and
connectors and for fiber modules.
Do not disconnect test cord 1 from the source. If you
do, you must set the reference again to make sure the
loss measurements are reliable.
For singlemode links with APC connectors, you must
use Type A test cords and adapters. Type A
components correctly align the angles in the
connectors.
5
On the source, select these settings:
SCAN ALL: ON
MODE: 2 kHz does not show
14
for 2.5 seconds
to put the cursor next to
.
Caution
or
to decrease
, then