MultiFiber Pro Getting Started Guide
To set the reference
For correct measurements, connect only APC
connectors to the singlemode sources. Connect only
non-APC connectors to the multimode source.
For singlemode links with APC connectors, you must
use Type A test cords and adapters. Type A
components correctly align the angles in the
connectors.
1
Clean the connectors on the meter, source, and a test cord.
2
Turn on the meter and source and let them warm up for 10
minutes. Allow more time if the equipment has been stored
above or below ambient temperature.
3
Make connections to set the reference. The figure on page
13 shows reference connections for links with unpinned
connectors. The Users Manual shows connections for links
with other types of connectors and for fiber modules.
4
On the source, hold down
setup menu, then select these settings:
 SCAN ALL: ON
 MODE: 2 kHz does not show
5
On the meter, press
6
If necessary, press
measurement for each channel.
7
To save the reference measurements, press
meter saves the measurements and the display briefly

shows
measurement mode.
12
to make SET REF show.
CHANNEL to see the power
, OK and REF. Then the meter goes into loss
Caution
for 2.5 seconds to see the
SAVE. The