Modality Type
DSA (Digital Subtraction
Radiography), including
C-Arm and Digital Fluoro
Phillips Medical Systems
(PMS) -
EasyVision and other PMS
modalities.
October 4, 2004
Table 1. Selection of Parameters by Modality Type
TFT Set
Table No.
(Contrast
Setting)*
VER713C0.w87
(Alternative 1)
VER693C0.w87
(Alternative 2)
PMSV3K.w87
Image
Density
(Dmax)*
6, 7, 9
2.9 - 3.1
6, 7, 9
2.9 - 3.1
9
3.0, 3.1
7F3319
Optimizing Image Quality
Smooth-
Note
ing Type
15
VER713C0 provides
brighter image
background than
VER693C0. Lower Table
numbers have less
mid-tone contrast.
Higher have more.
15
VER693C0 provides
darker image
background than
VER713C0. Lower Table
numbers have less
mid-tone contrast.
Higher have more.
15
Check with the PMS
service engineer. On
some PMS modalities,
image quality
calibration must be run
from the PMS control
console.