Sartorius ED224S-0CE Operating Instructions Manual page 17

Sartorius extend series; sartorius gem series; gold extend series. electronic analytical and precision balances and precious metal scales
Table of Contents

Advertisement

External Calibration
Parameters (changes in factory settings):
SETUP: Bal.Scal.: Cal.-Adj.: Cal.Ext. (menu code 1.1. 9. 1)
The required calibration weight is configured at the factory (see "Specifications")
Step
1. Tare the balance/scale
2. Start calibration.
Once you store the zero point
the required calibration weight
is prompted (flashing display)
3. Apply the prompted calibration
weight (in this example: 5000 g)
Weight too light:
a minus sign "–" is shown
Weight too heavy:
a plus sign "+" is shown
The display stops flashing
as soon as the weight value is
within the defined limit.
4. Calibration/adjustment executed;
then the calibration weight
is displayed
5. Remove the calibration weight
Key (or instruction)
w
Q
Display
0.0
g
Cal.Ext.
- 5000.0
g
5000.0
g
Cal.end
+ 5000.0
g
0.0
g
17