PCE Health and Fitness PCE-ATP 1 Manual page 22

Test instrument
Table of Contents

Advertisement

3. Installation: After sampling, please hold the joint port of the
ATP Quickswab and insert the swab tip back into the test tube.
(The end face of the test tube should be aligned with the lower
end face of the blue joint port).
4. Injection: Remove the cap of the ATP Quickswab, make sure it is
griped in upright position, and forcibly press down the spring
cap several times to the reagent fully flows into the bottom of the
test tube and submerge the swab tip.
5. Mixing: Hold the upper spring cap of the ATP Quickswab and
swing 30 degrees to the left and right (for five seconds), allowing
the reagent to fully react with the sample.
6. Insertion: Insert the ATP Quickswab into the test camber of
detector which is on the To-be-Tested interface, close the top
cover and start the test.
Fig1.3-1 Unfreezing
Fig1.3-2 Sampling
9