3
Instrument Features
3.3

Specifications

Dimensions (width x depth x height):
Weight:
Nominal supply voltages:
Nominal frequency:
Power draw:
Fuses:
IEC 1010 classification:
Overvoltage category:
Pollution degree:
International protection class:
A-weighted noise level:
Performance parameters
Slide throughput
Loading capacity:
Capacity of one specimen slide
holder:
Overall number of stations:
Reagent stations:
Volume of reagent vessels:
Overall number fo water stations:
Number of oven chambers:
Temperature of oven chamber:
Incubation time setting:
Overall number of loading / unloading
stations:
Capacity of nonvolatile storage:
Integration:
Fresh water connection
Hose material:
Hose length:
Connecting piece:
Inner diameter:
Outer diameter:
Internal pressure:
Required flow rate:
18
109 cm x 67 cm x 51 cm
65 kg
100-120V AC ± 10 % or 230-240V AC ± 10 %
50-60 Hz
650 VA
5 x 20 mm, VDE / UL listed
100-120 V: quick acting fuse F 6.30 A L250 VAC
230-240 V; quick acting fuse F 3.15 A L250 VAC
Protection class 1
II
2
IP20
≤ 70 dB (A)
depending on selected protocol:
max. 600 slides /h
min. 200 slides /h
max. 11 specimen slide holders
30 slides
26
min. 18
450 ml
max. 5 (DI water optional)
1
Ambient temperature or +30 to +65 °C
0 s to 99 min 59 s
1 / 1
15 programs with up to 25 steps
Connection with CV5030 (coverslipper, optional)
PVC
2.5 m
G3/8
10 mm
16 mm
Min. 1 bar / max. 6 bar
Min. 12 l/ min