8 Maintenance

Our customer service representatives are always at your disposal for any questions you might have.
This service includes any support regarding the proper application of your infrared measuring system,
calibration or the solution to customer-specific solutions as well as repair.
In many cases your problems will be applications-specific and can possibly be solved over the
telephone. So, if you need to return equipment to us, please contact our Service Department before
doing so, see phone and fax numbers at the beginning of this document.

8.1 Troubleshooting Minor Problems

8.2 Automatic Error Indication

The automatic error indication (alarm output) shall warn the user and guarantee a secure output in the
event of a system error. In the first place, however, its task is to switch the system off in case of a faulty
setup or a defect in the sensing head or in the electronic circuits.
Please select the desired Alarm Mode, see <Advanced Setup> in the menu <Sensor> <Sensor Setup> of
the software.
TX
info@Raytek-Direct.com
Symptom
No Output
Erroneous Temperature
Erroneous Temperature
Erroneous Temperature
Erroneous Temperature
Temperature Fluctuates
Off
Normally Open
Normally Closed
Intern. Temp. n.c.
Intern. Temp. n.o.
Current output:
Symptom
Internal defect
Target temperature under range
Target temperature over range
Never rely exclusively on the automatic error indication when monitoring critical
heating processes. It is strongly recommended to take additional safety measures!
Possible Cause
Cable Disconnected
Check Cable Connections
Faulty Sensor Cable
Verify Cable's Integrity
Field of View Obstructed
Remove the Obstruction
Lens Dirty
Clean the Lens
Wrong Emissivity Setting
Correct the Setting
Wrong Signal Processing
Correct Peak, Valley, or Average Settings
Tab. 3: Troubleshooting
no alarm function
open output within the alarm thresholds
closed output within the alarm thresholds
open, if internal sensor temperature is exceeded
closed if internal sensor temperature is exceeded
Rev. O7 07/2017
Maintenance
Solution
4 – 20 mA Output
21 mA
4 mA
20 mA