6.10 Protective Window

Protective windows can be used to protect the sensor's optic against dust and other contaminations. For
sensors with plastic lenses the use of a protective window combined with the air purge is strongly
recommended.
The following table provides an overview of the available protective windows recommended for a LT,
MT, HT, G5 and P7 model. All protective windows are also available in stainless steel (...S).
For special requirements, please ask our Sales Department about our range of special protective
windows.
Models delivered with mounted protective windows are preset as follows:
Basic model
Smart model
Determination of transmissivity of an unknown protective window:
If transmissivity of the measuring screen is not indicated on the data sheet, you can also determine the
transmissivity yourself. Please proceed as follows:
1.
2.
3.
For more information regarding the mounting of a protective window, see section 8.4
Protective Window
TX
info@Raytek-Direct.com
Order number
XXXTXACTWL
XXXTXACTWLF1
XXXTXACTWM
XXXTXACTWH
XXXTXACTWGP
Basic model has no adjustable parameter for the transmissivity. The influence of
transmissivity is reflected by a resulting emissivity value. E.g. transmissivity of
protective window is 0.88, emissivity of measured object is 0.95, resulting emissivity
value is set with the two rotary switches to 0.84 (0.88 * 0.95 = 0.84)
Transmissivity for the protective window is set in the software directly. The emissivity
value is not changed.
Measure the temperature of the target object with the sensing head, without using the protective
window. Note correct setting of emissivity. With the basic model, emissivity is set with the two
rotary switches. For the Smart model, the emissivity is entered in the software and the value for
"Transmissivity" is specified as 1.00.
Insert the protective window in the sensing head.
If you use the basic model of the sensing head, adjust the emissivity until the same temperature is
displayed, as was determined without the protective window.
If the smart model is used, adjust the transmissivity in the software until the same temperature is
displayed, as it was determined without the protective window.
on page 40.
Model
Designation
LT
none
LT
none (stainless steel)
MT
4 red dots
HT
3 red dots
G5, P7
2 red dots
Table 2: Protective Windows
Rev. O7 07/2017
Accessories
Material
0.65
Amtir
0.05
0.75
Polyethylene foil
0.05
(non-poisonous, non-fragile)
0.88
Sapphire
0.05
0.88
Glass
0.05
0.93
Calcium Fluorid
0.05
1.888.610.7664