8.4.2 Models produced before May 1999

Attention! If you own a model produced before May 1999 and where the protective window looks
similar to the one shown on the photographs below, refer to the following description.
Make sure to use the appropriate protective window for the spectral range of your sensor model, for
more information see section 6.10
The window material is placed in a metal ring which is enclosed with a rubber gasket. This rubber
gasket keeps the protective window in the sensor. (Models produced later than May 1999 have a thread
instead - check this, before you order spare parts!)
Replace the protective window using the plastic levers supplied with the spare window. Place the
levers' blades beneath the projection of the protective window. Now gently lever the protective window
from its mount.
TX
info@Raytek-Direct.com
Figure 34: Replacing the Protective Window
Protective Window
Figure 35: Replacing the Protective Window
Rev. O7 07/2017
Maintenance
on page 35.
Sensor
Window
Mounting Tool