Glp-Compliant Print - Scale House GAI420 Operation Manual

Gai series high-precision tuning fork electronic balance
Table of Contents

Advertisement

9.2 GLP-compliant print

At calibration, the AJH-CE Series enables printing in a format compliant with GLP. Set the function to the GLP print
support ([0.GLP 1]) side, then connect the balance to a printer or other equipment. When calibration is complete, the
balance prints out the format illustrated below.
Since only items are printed, the data should be entered manually by the person who conducted calibration.
※Reference: "4.2 Description of Functions" (Page 13)
"9.2 Calibration of AJH-CE Series" (Page 29)
GLP-Compliant Print Form
**CALI BRATI ON**
**CALI BRATI ON**
←見出し
M ODEL:
MO DEL:
←はかり の型名
S/N:
S/N:
←はかり の製造番号
I D:
I D:
←I D番号
DATE:
DATA:
←校正し た 日付
TI M E:
TI ME:
←校正し た 時刻
*CAL. END
*CAL. END
←校正終了
NAM E:
NAME:
←担当者署名
***************
***************
※To operate with CSP-160 printer, set On in DIP switch 3(Printing controlled by scale) and OFF in DIP switch 7 (disable
manual printing).
※For more information on the printer, refer to the printer's operation manual.
Caution:
Printing is done only when calibration ends normally. The balance will not print if calibration ends abnormally with
[3-Err], [4-Err], or [StOP] being issued.
Title
Model of balance
Serial No. of balance
ID No.
Date of calibration
Time of calibration
End of calibration
Signature of person in charge
33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gai620Gai2200Gai4200

Table of Contents