Limit Function; Limit Function Setup - Scale House GAI420 Operation Manual

Gai series high-precision tuning fork electronic balance
Table of Contents

Advertisement

The limit function judges measurements according to a limit value saved in the balance.
The function shows the judgement result by displaying the [
(insufficient). This function is very useful when discriminating between conforming and nonconforming articles. It is
also useful when measuring a given constant quantity consecutively, in conjunction with a range of reference weights
defined by upper- and lower-limit values.
This function can be used in weight mode, counting mode, or percentage mode.
Limit value input methods
Either of the following two methods can be used in the different modes:
(1)
Actual quantity setup method ........ An actual sample is loaded on the balance and its weight saved as the limit
(2)
Numeric value setup method ......... The limit value is entered with a key stroke.
※The limit values entered are held in memory, even when the balance is powered down.
※The respective limit values for weight mode, counting mode, and percentage mode are set up independently.
Indication of judgement result
The [
] mark lights up as either HI, OK, or LO on the left side of the display, indicating the result of judgement.
Judgement Results
Upper-limit value < measurement
HI (excessive)
value
Upper-limit value  measurement
OK (appropriate)
value  lower-limit value
Lower-limit value > Measurement
LO (insufficient)
value

8.1 Limit Function Setup

1
Invoking the function
Continuous
pressing
26

8. Limit Function

value.
Upper/lower-limit setting
Key released
] mark as either HI (excessive), OK (appropriate), or LO
One-point setting
No indication
Limit value  measurement value
Limit value > Measurement value
Press and hold down the
key.
Release the key when [Func] is
displayed.
The display changes to the function
setup and the first item is displayed.
Function

Hide quick links: