Increasing The Counting Accuracy (Memory Update Method) - Scale House GAI420 Operation Manual

Gai series high-precision tuning fork electronic balance
Table of Contents

Advertisement

6.2 Increasing the Counting Accuracy (Memory Update Method)

※This procedure is the same as the sampling procedure described on the previous page up, to the point at which the
sampling number is changed.
1
Loading samples
2
Saving the samples
3
Adding samples
4
Saving the samples
5
Repeating the addition and saving of samples
6
Terminating the sampling
10 pieces saved
20 pieces saved
125 pieces saved
Load the number of samples displayed.
Count the samples precisely and load
them on to the center of the
measurement pan.
Press the
Set
key.
The unit weight of the samples is saved,
and the display changes to a flashing
value. This flashing value denotes that
the Memory Update Method is currently
active.
Add approximately an equal number of
samples as the number currently
displayed. Add them gradually, in
groups of approximately the same
number, until the number of samples is
approximately double. The additional
samples need not be counted.
Press the
Set
key.
After the unit weight of the samples has
been saved, the display begins to flash,
indicating that the Memory Update
Method is still running.
Repeat steps 3 and 4. The eventual
total number of samples to be saved
should be approximately 1/5 to 1/2 of
the number of samples to be measured.
Press the
Memory
key.
The balance saves the unit weight and
returns you to measurement mode.
23

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Gai620Gai2200Gai4200

Table of Contents