5 CEPHALOMETRIC EXPOSURE
30
Planmeca ProMax Cephalostat
If you need to adjust the preset exposure values for the
current exposure:
1.
Select the kV / mA field.
2.
Use the minus or plus buttons to set the exposure
values you wish to use. To improve the image
contrast, reduce the kV value. To reduce the
radiation dose, reduce the mA value.
3.
Select the green check mark button.
1
NOTE
To adjust the preset exposure values permanently, select
Settings > Program > 2100 Programs > 2140
2D Cephalostat. Refer to section "Program settings" in the
User's Manual of the X-ray unit for details.
2
3
User's Manual