5.
Press the release lever on the left ear post holder and
slide the ear posts and nasal positioner out as far as
they will go.
6.
Adjust the height of the cephalostat by pressing a
height adjusting button until the positioning cones at
the ends of the ear posts are level with the patient's
ears.
7.
Position the patient between the ear posts so that the
patient faces the nasal positioner.
8.
Press the release lever on the left ear post holder and
very carefully slide the positioning cones into the
patient's ears.
9.
Slide the nasal positioner until it touches the patient's
nasion.
10. Adjust the angle of the patient's head by sliding the
nasal positioner up or down until the Frankfort plane
is horizontal.
Planmeca ProMax Cephalostat
5 CEPHALOMETRIC EXPOSURE