10.2 What is WB
Wet Bulb Temperature -- WB
Wet-bulb temperature is measured with a wet-bulb thermometer, which is a regular thermometer with a wet
muslin wick cover.
To calculate wet-bulb temperature without using a wet-bulb thermometer, existing dry-bulb temperature and
relative temperature are used on the aspirated psychrometer humidity table that is JIS standard Z8806
compliant. In this manual; however, we uses Newtonian approximation based on the assumption of a
temperature measured on a wet-bulb thermometer being lower than a dry-bulb thermometer.
ln(e
)=-6096.9385×T
tw
+1.673952×(10
A= e
tw
f'(tw)=4030.183/((235+tw)
tw1=tw-(A-P×(t-tw)/2/755-E×U/100)/f'(tw)
tw: Wet bulb temperature (℃)
e
: Saturated water vapor pressure at tw (Pa)
tw
T
: Absolute temperature (k)=(tw+273.15)
w
P: Barometric pressure (Pa)
E: Saturated water vapor pressure at t (Pa)
U: Relative humidity
T: Dry bulb temperature (℃)
10.3 What is AH
Absolute Humidity -- AH
Absolute humidity represents an amount of water vapor contents per 1kg of dry air.
To calculate absolute humidity, apply temperature and relative humidity on below formula.
ln(e
)=-6096.9385×T
w
+1.673952×(10
e=U/100×eW
D(g/m3)=0.794×(10
e
: Saturated water vapor pressure (Pa)
w
T: Absolute temperature (K)=t(℃)+273.15
t: Dry bulb temperature (℃)
e: Water vapor pressure (Pa)
U: Relative humidity
D: Absolute humidity (g/m3)
netzerotools.com
-1
+21.2409642-2.711193×(10
w
-5
2
) × T
+2.433502×ln(T
w
2
)×A+P/2/755
-1
+21.2409642-2.711193×(10
-5
2
) ×T
+2.433502×ln(T)
-2
) ×e/(1+0.00366×t)
netzerotools.com
-2
) × T
w
)
w
-2
) ×T
46