YOKOGAWA TA520 User Manual page 94

Time interval analyzer gp-ib interface
Table of Contents

Advertisement

Appendix 2 Error Messages
Error in communication Query (400 to 499)
Code Message
410
Query INTERRUPTED
420
Query UNTERMINATED
430
Query DEADLOCKED
440
Query UNTERMINATED after
indefinite response
Error in Execution (600 to 899)
See User's Manual IM704310-01E.
Error in System Operation (912 to 915)
Code Message
912
Fatal error in Communication-
driver
914
A communication time-out has
occurred
Other (350)
Code Message
350
Queue overflow
App-4
Action
The transmission of the response was aborted.
Check the transmission/reception order.
No response to be transmitted.
Check the transmission/reception order.
Transmission/reception deadlock occurred.
Abort the transmission.
Limit the length of the program messages to 1024 bytes
including <PMT>.
The request order is not correct.
Do not enter a query after the
Action
Communication driver error occurred.
Servicing is required.
Communication timeout error occurred.
Make sure to receive the data within the timeout time.
The communication line may be malfunctioning.
Action
Read out the error queue.
Note
Code 350 indicates overflow of error queue. This code is returned as a response to the
query; it does not appear on the screen.
"STATus:ERRor?"
and
commands.
*IDN?
*OPT?
Reference Page
2-2
2-2
2-2
Reference Page
Reference Page
4-5
IM 704310-12E