YOKOGAWA TA520 User Manual page 73

Time interval analyzer gp-ib interface
Table of Contents

Advertisement

THIStogram:ISI:SPACe<x>
Function
Sets the space or queries the current setting.
Syntax
:THIStogram:ISI:SPACe<x> <NRf>
:THIStogram:ISI:SPACe<x>?
<NRf>=Window label
Example
:THIStogram:ISI:SPACe<x> 3
:THIStogram:ISI:SPACe<x>?
→:THIStogram:ISI:SPACe<x> 3
Description
When the trigger condition is SPACe or
STOMark, only SPACe1 can be specified. For
MTOSpace, only SPACe2 can be specified. For
STOSpace, both SPACe1 and SPACe2 can be
specified. For all other cases SPACe<x>
cannot be specified.
THIStogram:ISI[:STATe]
Function
Turns ON/OFF the inter-symbol interference
analysis function or queries the current setting.
Syntax
:THIStogram:ISI[:STATe] <Boolean>
:THIStogram:ISI:STATe?
Example
:THISTOGRAM:ISI:STATE ON
:THISTOGRAM:ISI:STATE?
→:THISTOGRAM:ISI:STATE ON
THIStogram:ISI:TARGet
Function
Sets the item to be analyzed or queries the
current setting.
Syntax
:THIStogram:ISI:TARGet {PREVious|NEXT|
BOTH}
:THIStogram:ISI:TARGet?
Example
:THISTOGRAM:ISI:TARGET PREVIOUS
:THISTOGRAM:ISI:TARGET?
→:THISTOGRAM:ISI:TARGET PREVIOUS
Description
When the trigger mode is SINGle, PREVious,
NEXT, or BOTH can be specified. For
COMBination, only PREVious or NEXT can be
specified. For BETWEEN, nothing can be
specified.
THIStogram:ISI:TRIGger
Function
Sets the trigger condition or queries the current
setting.
Syntax
:THIStogram:ISI:TRIGger
{MARK|SPACe|MTOSpace|STOMark|MTOMark|
STOSpace}
:THIStogram:ISI:TRIGger?
Example
:THISTOGRAM:ISI:TRIGGER MARK
:THISTOGRAM:ISI:TRIGGER?
→:THISTOGRAM:ISI:TRIGGER MARK
Description
When the trigger mode is SINGle, MARK or
SPACE can be specified. For COMBination,
MTOSpace or STOMark can be specified. For
BETWEEN, MTOMark or STOSpace can be
specified.
IM 704310-12E
THIStogram:ISI:TVALue
Function
Sets the value of the constant T or queries the
current setting.
Syntax
:THIStogram:ISI:TVALue <Time>
:THIStogram:ISI:TVALue?
<Time>=1 ns to 250 ns (in 25 ps steps)
Example
:THISTOGRAM:ISI:TVALUE 100ns
:THISTOGRAM:ISI:TVALUE?
→:HHISTOGRAM:SINGLE:TVALUE 0.100000E-06
THIStogram:ISI:UPDate
Function
Changes the window setting based on the
constant T.
Syntax
:THIStogram:ISI:UPDate
Example
:THISTOGRAM:ISI:UPDATE
THIStogram:ISI:WINDow<x>?
Function
Queries all settings related to each window.
Syntax
:THIStogram:ISI:WINDow<x>?
Example
:THISTOGRAM:ISI:WINDOW1?
→:THISTOGRAM:ISI:WINDOW1:LABEL 3;
HORIZONTAL:CENTER 150.000E-09;
SPAN 0.3E-06;:THISTOGRAM:ISI:WINDOW1:
MARKER:LEFT 100.000E-09;
RIGHTT 200.000E-09
THIStogram:ISI:WINDow<x>:HORizontal?
Function
Queries all settings related to the horizontal
axis of each window.
Syntax
:THIStogram:ISI:WINDow<x>:HORizontal?
Example
:THISTOGRAM:ISI:WINDOW1:HORIZONTAL:
CENTER?
→:THISTOGRAM:ISI:WINDOW1:HORIZONTAL:
CENTER 150.000E-09;SPAN 0.3E-06
THIStogram:ISI:WINDow<x>:HORizontal:CENTer
Function
Sets the center value of the horizontal axis or
queries the current setting.
Syntax
:THIStogram:ISI:WINDow<x>:HORizontal:
CENTer <Time>
:THIStogram:ISI:WINDow<x>:HORizontal:
CENTer?
<Time>=-50 ns to 3.2 µs (in 25 ps steps)
Example
:THISTOGRAM:ISI:WINDOW1:HORIZONTAL:
CENTER 150ns
:THISTOGRAM:ISI:WINDOW1:HORIZONTAL:
CENTER?
→:THISTOGRAM:ISI:WINDOW1:HORIZONTAL:
CENTER 150.000E-09
3.21 THIStogram Group
3-53
3

Advertisement

Table of Contents
loading

Table of Contents