YOKOGAWA TA520 User Manual page 35

Time interval analyzer gp-ib interface
Table of Contents

Advertisement

CALCulation:PARameter:JITTer
Function
Turns ON/OFF the calculation of the jitter value
(σ/T) during the histogram display or queries
the current setting.
Syntax
:CALCulation:PARameter:JITTer <Boolean>
:CALCulation:PARameter:JITTer?
Example
:CALCULATION:PARAMETER:JITTER ON
:CALCULATION:PARAMETER:JITTER?
→:CALCULATION:PARAMETER:JITTER 1
CALCulation:PARameter:MAXimum
Function
Turns ON/OFF the calculation of the maximum
value during the histogram display or queries
the current setting.
Syntax
:CALCulation:PARameter:MAXimum <Boolean>
:CALCulation:PARameter:MAXimum?
Example
:CALCULATION:PARAMETER:MAXIMUM ON
:CALCULATION:PARAMETER:MAXIMUM?
→:CALCULATION:PARAMETER:MAXIMUM 1
CALCulation:PARameter:MEDian
Function
Turns ON/OFF the calculation of the median
value during the histogram display or queries
the current setting.
Syntax
:CALCulation:PARameter:MEDian <Boolean>
:CALCulation:PARameter:MEDian?
Example
:CALCULATION:PARAMETER:MEDIAN ON
:CALCULATION:PARAMETER:MEDIAN?
→:CALCULATION:PARAMETER:MEDIAN 1
CALCulation:PARameter:MELE
Function
Turns ON/OFF the calculation of the MELE
value during the histogram display or queries
the current setting.
Syntax
:CALCulation:PARameter:MELE <Boolean>
:CALCulation:PARameter:MELE?
Example
:CALCULATION:PARAMETER:MELE ON
:CALCULATION:PARAMETER:MELE?
→:CALCULATION:PARAMETER:MELE 1
CALCulation:PARameter:MINimum
Function
Turns ON/OFF the calculation of the minimum
value during the histogram display or queries
the current setting.
Syntax
:CALCulation:PARameter:MINimum <Boolean>
:CALCulation:PARameter:MINimum?
Example
:CALCULATION:PARAMETER:MINIMUM ON
:CALCULATION:PARAMETER:MINIMUM?
→:CALCULATION:PARAMETER:MINIMUM 1
IM 704310-12E
CALCulation:PARameter:MODE
Function
Turns ON/OFF the calculation of the most
frequent value during the histogram display or
queries the current setting.
Syntax
:CALCulation:PARameter:MODE <Boolean>
:CALCulation:PARameter:MODE?
Example
:CALCULATION:PARAMETER:MODE ON
:CALCULATION:PARAMETER:MODE?
→:CALCULATION:PARAMETER:MODE 1
CALCulation:PARameter:PTOPeak
Function
Turns ON/OFF the calculation of the P-P value
during the histogram display or queries the
current setting.
Syntax
:CALCulation:PARameter:PTOPeak <Boolean>
:CALCulation:PARameter:PTOPeak?
Example
:CALCULATION:PARAMETER:PTOPEAK ON
:CALCULATION:PARAMETER:PTOPEAK?
→:CALCULATION:PARAMETER:PTOPEAK 1
CALCulation:PARameter:SDEViation
Function
Turns ON/OFF the calculation of the standard
deviation (σ) during the histogram display or
queries the current setting.
Syntax
:CALCulation:PARameter:SDEViation
<Boolean>
:CALCulation:PARameter:SDEViation?
Example
:CALCULATION:PARAMETER:SDEVIATION ON
:CALCULATION:PARAMETER:SDEVIATION?
→:CALCULATION:PARAMETER:SDEVIATION 1
CALCulation:PARameter:TAVerage
Function
Turns ON/OFF the calculation of the average
value during the time variation display or
queries the current setting.
Syntax
:CALCulation:PARameter:TAVerage
<Boolean>
:CALCulation:PARameter:TAVerage?
Example
:CALCULATION:PARAMETER:TAVERAGE ON
:CALCULATION:PARAMETER:TAVERAGE?
→:CALCULATION:PARAMETER:TAVERAGE 1
CALCulation:PARameter:TJITter
Function
Turns ON/OFF the calculation of the jitter value
(P-P/average value) during the time variation
display or queries the current setting.
Syntax
:CALCulation:PARameter:TJITter <Boolean>
:CALCulation:PARameter:TJITter?
Example
:CALCULATION:PARAMETER:TJITTER ON
:CALCULATION:PARAMETER:TJITTER?
→:CALCULATION:PARAMETER:TJITTER 1
3.3 CALCulation Group
3-15
3

Advertisement

Table of Contents
loading