Pulse Output, Change Active/Passive - ABB FSM4000 Operating Instructions Manual

Fieldit electromagnetic flowmeter with ac field technology converter model s4
Hide thumbs Also See for FSM4000:
Table of Contents

Advertisement

9.9
The factory setting for the pulse output is noted on the type tag on the converter can be obtained from the
Ordering Information (Model Number). Use the following procedure if a change is required:
Fig. 22:
Fig. 23:
Fig. 24:
54
S4

Pulse Output, Change Active/Passive

Turn off supply power
After a wait time of 40 minutes, unscrew the housing cover of the field mount housing
Loosen the two screws on the opposite sides and remove the converter module (see Fig. 22)
The jumper for changing the settings may be found on the underside of the module (Fig. 22). Install
jumper in the desired location:
– Active
= outside see Fig. 23
– Passive = inside see Fig. 24
Reinstall the converter in the housing and tighten the screws
Close the cover and tighten the screws
Mounting screw
Converter Module
Jumper Position „Active"
Jumper Position „Passive"
9 Maintenance
Mounting screw
Jumper location
D184B122U02