Data Entry - ABB FSM4000 Operating Instructions Manual

Fieldit electromagnetic flowmeter with ac field technology converter model s4
Hide thumbs Also See for FSM4000:
Table of Contents

Advertisement

8

Data Entry

8.1
The data can be entered in a number of languages using the three keys on the converter. After the power is
turned on, the instrument automatically conducts a number of selftest routines. After they have been
successfully completed the standard display appears (process information). The variables to be displayed
can be selected by the user.
Fig. 13:
Fig. 14:
Data can be entered using the Magnet Stick without opening the housing cover.
Important!
If a converter design is used with the tropicalization option – high humidity –, data entry should use the Mag-
net Stick exclusively.
Warning!
STOP
When the housing cover is removed EMC and personnel contact protection is no longer provided.
The values can be displayed in a four line format, or for improved readability at the meter location, in a two
line format with larger characters. The values to be displayed can be individually selected in the submenu
Display. The contrast setting in the submenu Display can be used to adjust the display for the ambient light
conditions. During data entry the converter remains online, i.e., the current and pulse outputs continue to
indicate the actual flowrate values. During a power outage or when the power is turned off, all parameter
settings and totalizer values are securely stored in a FRAM.
22
S4
General Description
DATA
Function
Converter Keypad and Display
Magnet Stick Operation
FSM4000
N
S
N
S
N
S
STEP
C/CE
Direct data entry using
ENTER
the keypad
Magnet Stick
Operation without opening
the housing
8 Data Entry
Data entry using a
Magnet Stick without
opening the housing
D184B122U02

Advertisement

Table of Contents
loading

This manual is also suitable for:

Se21 seriesSe41f

Table of Contents