ABB AX413 User Manual page 63

Analyzeit single and dual input analyzers for high level conductivity
Hide thumbs Also See for AX413:
Table of Contents

Advertisement

...Analog Outputs
Maximum load resistance
750R ohms at 20mA
Configuration
Can be assigned to either measured variable or
either sample temperature
Digital Communications
Communications
Profibus DP (with option board fitted)
Control Function – AX430 Only
Controller Type
P, PI, PID (Configurable)
Control Outputs
Analog
Current output control (0 to 100%)
Time proportioning cycle time
1.0 to 300.0s, programmable in increments of 0.1s
Pulse frequency
1 to 120 pulses per minute, programmable in increments of
1 pulse per minute
Controller action
Direct or reverse
Proportional band
0.1 to 999.9%, programmable in increments of 0.1%
Integral action time (Integral reset)
1 to 7200s, programmable in increments of 1s (0 = Off)
Derivative
0.1 to 999.9s in increments of 0.1s,
available only for single set point control
Auto/Manual
User-programmable
Access to Functions
Direct keypad access
Measurement, maintenance, configuration,
diagnostics or service functions
Performed without external equipment or internal jumpers
Mechanical Data
Wall-/Pipe-mount versions
IP66/NEMA4X
Dimensions 192mm high x 230mm wide x 94mm deep
(7.56 in. high x 9.06 in. wide x 3.7 in. deep)
Weight 1kg (2.2 lb)
Cable Entry Types
Standard
5 or 7 x M20 cable glands
N. American
7 x knockouts suitable for
Panel-mount versions
IP66/NEMA4X (front only)
Dimensions 96mm x 96mm x 162mm deep
(3.78 in. x 3.78 in. x 6.38 in. deep)
Weight 0.6kg (1.32 lb)
Power Supply
Voltage requirements
85 to 265V AC 50/60 Hz
24V AC or 12 to 30V DC (optional)
Power consumption
<10VA
Insulation
Mains to earth (line to ground) 2kV RMS
Environmental Data
Operating temperature limits
–20 to 65°C (–4 to 149°F)
Storage temperature limits
–25 to 75°C (–13 to 167°F)
Operating humidity limits
Up to 95%RH non condensing
EMC
Emissions and immunity
Meets requirements of:
Hazardous area approvals
CENELEC ATEX IIG EEx n IIC T4
FM non-incendive Class I Div. 2 Groups A to D
CSA non-incendive Class I Div. 2 Groups A to D
Safety
General safety
EN61010-1
Overvoltage Class II on inputs and outputs
Pollution category 2
Languages
Languages configurable:
1
/2 in. Hubble gland
EN61326 (for an industrial environment)
EN50081-2
EN50082-2
English
Italian
French
Spanish
German
SPECIFICATION
Pending
Pending
Pending
SS/AX4CO4 Issue 1
61

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax430Ax436Ax433Ax438

Table of Contents