...5.3 Configure Sensors
A: Temp.Comp.
set to
Linear
2. 0 0
%/C
A: Temp.Coeff.
10
Secs
A: Filter Time
Config. Sensor A
A: Temp.Comp.
set to
User
Edit
Yes
-----
No
User Temp Comp
No
10
Secs
A: Filter Time
Config. Sensor A
Temperature Coefficient
Enter the temperature coefficient ( x 100) of the solution (0.01 to 5.0%/°C). If unknown,
the temperature coefficient () of the solution must be calculated – see Appendix A1.1.
If the value has not yet been calculated, set it to 2%/°C provisionally.
Filter Time
Set the required filter time, between 1 and 60 seconds in 1 second increments.
Sensor B configuration (dual input analyzers only) is identical to Sensor A
Config. Sensor B
configuration.
See Section 5.4.
CONFIG. ALARMS
Edit User Defined Temperature Compensation
Select
to manually define a 6-point temperature compensation curve, otherwise
Yes
select
.
No
Yes
Continued on next page.
A: Enter Point 1
Filter Time
Set the required filter time, between 1 and 60 seconds in 1 second increments.
Sensor B configuration (dual input analyzers only) is identical to Sensor A
Config. Sensor B
configuration.
See Section 5.4.
CONFIG. ALARMS
5 PROGRAMMING...
19