ABB AX413 User Manual page 25

Analyzeit single and dual input analyzers for high level conductivity
Hide thumbs Also See for AX413:
Table of Contents

Advertisement

...5.3 Configure Sensors
Define
ppb
ppm
-----
%
A: User Units
%, ppb or ppm
Define
19. 9 9
aaaaaa
A: Define
19. 9 9
%
A: Decimal Place
10
Secs
A: Filter Time
Config. Sensor A
User Units
Select the required concentration units:
%
– concentration as a percentage
– parts-per-million
ppm
ppb
– parts-per-billion
Define
– User defined units
Define User Units
Use the
and
keys to select a character at the cursor position. Use the
enter the character and move the cursor.
Press the
key when complete to advance to the next frame.
Decimal Place
Set the position of the decimal point.
Filter Time
Set the required filter time, between 1 and 60 seconds in 1 second increments.
Sensor B configuration (dual input analyzers only) is identical to Sensor A
Config. Sensor B
configuration.
See Section 5.4.
CONFIG. ALARMS
5 PROGRAMMING...
key to
23

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax430Ax436Ax433Ax438

Table of Contents