...7.3 Factory Settings
25. 0 0
Ohms
25. 0 0
Calib
A: Res 25
ohm
2. 5 00
Ohms
2. 5 00
Calib
A: Res 2.5 ohm
0. 5 00
Ohms
0. 5 00
Calib
A: Res 0.5 ohm
1000
Ohms
1000
Calib
A:T.Zero
(1K0)
1500
Ohms
1500
Calib
A:T.Span
(1K5)
Resistance Span (25R ohms)
Set the cell simulator to 25R ohms
The display advances automatically to the next step once a stable and valid value is
recorded.
Resistance Span (2.5R ohms)
Set the cell simulator to 2.5R ohms
The display advances automatically to the next step once a stable and valid value is
recorded.
Resistance Span (500R ohms)
Set the cell simulator to 500R ohms
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Zero (1000R ohms)
Set the temperature simulator to 1000R ohms
The display advances automatically to the next step once a stable and valid value is
recorded.
Temperature Span (1K5)
Set the temperature simulator to 1500R ohms
The display advances automatically to the next step once a stable and valid value is
recorded.
Continued on next page.
A: Ref. Checking
7 CALIBRATION...