ABB AX413 User Manual page 24

Analyzeit single and dual input analyzers for high level conductivity
Hide thumbs Also See for AX413:
Table of Contents

Advertisement

...5 PROGRAMMING
...5.3 Configure Sensors
Concentration
set to
Yes
0. 0
Deg.C
1. 7 9
Ratio
A: Enter point 1
0. 0
Deg.C
1. 7 9
Ratio
A: Enter point 1
125. 0
Deg.C
0. 3 0
Ratio
A: Enter point 6
0. 0
mS/cm
0. 0 0
%
A: Enter point 1
0. 0
mS/cm
0. 0 0
%
A: Enter point 1
200. 0
mS/cm
20. 0
%
A: Enter point 6
22
Temperature Value – Point 1
(or
) and
Deg. C
Deg. F
Adjust
and
keys to adjust the displayed reading to the required temperature value for point 1
on the curve.
Ratio Value – Point 1
and
are shown alternately on the center display line. Use the
Ratio
Adjust
to adjust the displayed reading to the required ratio value for point 1 on the curve.
Temperature and Ratio Values – Points 2 to 6
Enter the temperature and ratio values for the remaining points in the user-defined
temperature compensation curve.
Conductivity Value – Point 1
and
are shown alternately on the upper display line. Use the
mS/cm
Adjust
keys to adjust the displayed reading to the required conductivity value for point 1 on the
curve.
Concentration Value – Point 1
and
are shown alternately on the center display line. Use the
%
Adjust
adjust the displayed reading to the required concentration value for point 1 on the curve.
Conductivity and Concentration Values – Points 2 to 6
Enter the conductivity and concentration values for the remaining points in the user-
defined concentration compensation curve.
Continued on next page.
A: User Units
are shown alternately on the upper display line. Use the
and
keys
and
and
keys to

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax430Ax436Ax433Ax438

Table of Contents