ABB AX413 User Manual page 29

Analyzeit single and dual input analyzers for high level conductivity
Hide thumbs Also See for AX413:
Table of Contents

Advertisement

...5.4 Configure Alarms
...5.4.2 Configure CIP Interface Alarm
No
-----
Yes
Dosing
No
Yes
10
Mins
Dosing Time
No
Yes
Bandwidth
Config. Alarm 1
No
Yes
10
%
Dev. Setpoint
50
Secs
Delay Time
Config. Alarm 1
Dosing
Select
to enable dosing action, otherwise select
Yes
Dosing is inhibited until the initial charge sequence is complete and the measured
conductivity exceeds the Alarm 1 set point value.
Dosing is enabled when the measured conductivity falls below the Alarm 1 set point value
and continues for the dosing time period set below.
Dosing Time
Set the dosing period, between 1 and 10 minutes in 1 minute increments.
Bandwidth Alarm
Select
to enable the bandwidth alarm, otherwise select
Yes
The bandwidth alarm is activated if the measured conductivity either exceeds the Alarm 1
set point value plus the deviation set point value (see below), or falls below the Alarm 1 set
point value minus the deviation set point value.
If
Example:
A1: Setpoint
bandwidth alarm is activated if the process variable exceeds 1250mS/cm.
Deviation Set Point
Set the deviation set point, between 1 and 100% of the Alarm 1 set point value in 1%
increments.
Delay Time
If a bandwidth alarm condition occurs, activation of the relays and LEDs can be delayed for
the specified time period. If the alarm clears within the period, the alarm is not activated.
Set the required delay, between 0 and 60 seconds in 1 second increments.
Continued on next page.
Config. Alarm 2
See Section 5.5.
CONFIG. OUTPUTS
No
is set to
and
1000mS/cm
Dev. Setpoint
5 PROGRAMMING...
.
.
No
is set to
25%
, the
27