...5 PROGRAMMING
...5.3 Configure Sensors
User Temp Comp
set to
Yes
0. 0
Deg.C
1. 7 9
Ratio
A: Enter point 1
0. 0
Deg.C
1. 7 9
Ratio
A: Enter point 1
125. 0
Deg.C
0. 3 0
Ratio
A: Enter point 6
10
Secs
A: Filter Time
Config. Sensor A
20
Temperature Value – Point 1
(or
) and
Deg. C
Deg. F
Adjust
and
keys to adjust the displayed reading to the required temperature value for point 1
on the curve.
Ratio Value – Point 1
and
are shown alternately on the center display line. Use the
Ratio
Adjust
to adjust the displayed reading to the required ratio value for point 1 on the curve.
Temperature and Ratio Values – Points 2 to 6
Enter the temperature and ratio values for the remaining points in the user-defined
temperature compensation curve.
Filter Time
Set the required filter time, between 1 and 60 seconds in 1 second increments.
Sensor B configuration (dual input analyzers only) is identical to Sensor A
Config. Sensor B
configuration.
See Section 5.4.
CONFIG. ALARMS
are shown alternately on the upper display line. Use the
and
keys