5. Operation
5.3. Transmitter PV display
PV status
(blank)
BAD
UNC
Transmitter
Wireless Builder display
display
OUT SVC
OOS
CAL ERR
Calibration Error
CFG ERR
Configuration Error
LOW BAT
Low Battery
NO RADIO
Radio Interprocessor
Comm Error
BAD RADIO
Sensor Radio SPI
SPI
Communication Failure
BAD RADIO
EEPROM SPI
EEPROM
Communication Failure
RADIO WDT
WDT Reset Occurred
RESET
22
OneWireless XYR 6000 SmartCET Corrosion Transmitter User's Manual
Table 6 PV status
•
PV is normal – no action required
•
Possible calibration error – Clear calibration
AITB can not execute due to internal firmware state – Attempt cold restart of
•
device.
AITB can not execute due to hardware fault – Replace sensor board
•
•
Sensor failure – Check input connections
•
Sensor failure – Check bad probe
Warning: Input inaccurate due to uncertain input data integrity.
•
•
Warning: Input inaccurate due to input conversion limitations or resolution.
•
Warning: Input outside of characterized range. Value is estimated.
Table 7 Device status
Definition
All channels are out of
service.
Calibration Data Invalid
or could not be read.
Configuration Check
Error.
Battery Voltage Critically
Low
Radio Board is not
accessible.
Radio detected loss of
communication with
sensor board over the
inter-processor
communication link.
Radio EEPROM SPI
Communication failure
Radio Watch Dog
Timeout detected
Cause - Action
What to do
Restore mode to Auto in Wireless
Builder.
Use Cal Clear, Restore, or User
Calibrate.
Database is corrupted. Cold start
and reload configuration.
Replace batteries as soon as
possible. See page 47.
Restart both the radio and sensor.
If condition persists, replace
sensor module. See page 45.
Restart both the radio and sensor.
If condition persists, replace
sensor module. See page 45.
The radio will not be able to
perform firmware upgrades but will
operate normally using installed
code. Replace sensor module.
See page 45.
Restart both the radio and sensor.
If condition persists, replace
sensor module. See page 45.
Revision 3
6/24/08