6050H Calibration Bath
User's Guide
1.3
6
• Under filling the bath may reduce the bath performance and may possibly
damage the bath.
• When calibrating PRTs always follow correct calibration procedure and
calibrate from high temperatures to low temperatures with the appropriate
triple point of water checks. Never immerse a wet or cold PRT into a bath
filled with hot fluid. Severe damage to the PRT may result as well as per-
sonal injury to the calibration technician.
• This bath is not designed to be portable. Therefore, moving the bath once
it has been installed should be kept to a minimum. NEVER MOVE A
BATH THAT IS FULL OF "HOT" FLUID. This action could be ex-
tremely dangerous and could result in personal injury to the person mov-
ing the bath. If the bath is going to be placed in an area where it may need
to be moved frequently, Hart Scientific sells a special cart designed to ac-
commodate the bath. However, even with a cart the bath should not be
moved full of "hot" fluid. The fluid can splash causing injury or if the
bath and cart tip, the "hot" fluid could cause damage to the surrounding
area and personnel.
• If the bath must be moved, allow the salt to solidify and cool. Care must
be used when moving the bath with the fluid solidified in the bath. The
control probe can easily be damaged. Two people are required to safely
move the bath. One person should lift the bath by the tower, while the sec-
ond person lifts from underneath the bath tank.
• The control probe must be inserted through the lid into the bath fluid and
plugged into the socket at the back of the bath. DO NOT operate the bath
without the control probe properly inserted and attached. The bath will
not operate correctly without the control probe. Injury to operating per-
sonnel and permanent damage to the bath could occur.

Authorized Service Centers

Please contact one of the following authorized Service Centers to coordinate
service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services

Advertisement