Raytech WR14-R Instruction Manual page 138

Winding resistance meter 15a, 50a, 100a
Table of Contents

Advertisement

Erratic or erroneous readings:
Possible causes:
1. Test lead damaged or not connected.
2. Poor test lead connection.
3. Poor Mains power source supply. Check extension cords, plugs and cables carefully.
4. Test set attached to a "Live" load or high interference load.
5. Try to measure the primary winding in series to the secondary winding. See chapter F
Measuring high- and low-voltage-side in series on page 125.
Cannot switch off the WR14
Disconnect the power cable and press the Power switch for 15 seconds.
Instruction Manual Winding Resistance Meter
Type WR100 / WR50 / WR14
NOTE:
The WR series instruments are designed to be trouble free.
If problems or questions do arise please contact your nearest
representative or our service support group in Switzerland.
Page 138 of 139