IMMUNITY test
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
IEC 60601 test level
± 1 kV line(s) to line(s)
<5 % UT (>95 % dip in
UT) for 0,5 cycle
40 % UT (60 % dip in
UT) for 5 cycles
70 % UT (30 % dip in
UT) for 25 cycles
<5 % UT (>95 % dip in
UT) for 5 s
3 A/m
Note: UT is the a.c. mains voltage prior to application of the test level.
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic
environment specified below. The user of the Philips Avent breast pump
should assure that it is used in such an environment.
Compliance level
± 1 kV line(s) to line(s)
<5 % UT (>95 % dip in
UT) for 0,5 cycle
70 % UT (30 % dip in
UT) for 25 cycles
<5 % UT (>95 % dip in
UT) for 5 s
3 A/m
Electromagnetic
environment –
guidance
Mains power quality
should be that of a
typical residential or
hospital environment.
Mains power quality
should be that of a
typical residential or
hospital environment. If
the user of the Philips
Avent breast pump
requires continued
operation during power
mains interruptions, it is
recommended that the
Philips Avent breast
pump be powered from
an uninterruptible
power supply or a
battery.
Interruptible electric
supply can cause the
breast pump to turn off.
This is acceptable as
this will not lead to
unacceptable risks.
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical residential or
hospital environment.