Emissions test
RF emissions CISPR
11
RF emissions CISPR
11
Harmonic
emissions IEC
61000-3-2
Voltage
fluctuations/ flicker
emissions IEC
61000-3-3
IMMUNITY test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast
transient/burst IEC
61000-4-4
The electronic breast pump has no essential performance. Due to EMC
influences, like wireless home network devices, mobile phones or cordless
telephones, the breast pump can turn off or go into error mode. This will not
lead to unacceptable risks. To prevent interference, keep other electric
equipment at least 3.3 feet/1.0 m away from the breast pump during
expressing and do not stack with other electrical equipment. Cable length of
the adapter: 2.50 meter/ 8.2 feet.
Electromagnetic Compatibility (EMC, IEC 60601-1-2)
Declaration – electromagnetic emissions
The Philips Avent breast pump is intended for use in the electromagnetic
environment specified below. The user of the Philips Avent breast pump
should assure that it is used in such an environment.
Compliance
Electromagnetic environment – guidance
Group 1
The Philips Avent breast pump uses RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
Class B
Class A
The Philips Avent breast pump is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low voltage power supply
network that supplies buildings used for domestic purposes.
Complies
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic
environment specified below. The user of the Philips Avent breast pump
should assure that it is used in such an environment.
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
English
Electromagnetic
environment –
guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be at least 30 %.
Mains power quality
should be that of a
typical residential or
hospital environment.