On an InfiniiVision Oscilloscope

The N7040A/41A/42A probe is not automatically identified by the oscilloscope and therefore you
need to manually set the attenuation factor and units for the probe.
1
2
3
4
5
N7040A/1A/2A Rogowski AC Current Probes User Guide
Measure the current and observe the waveform on the oscilloscope. Remember that the actual
current will be greater or less than the oscilloscope waveform if the probe attenuation has not
been set accurately.
Press the channel key to which the current probe is attached to display the Channel <N> menu.
Press the Coupling softkey to set the oscilloscope's Input Coupling to AC to remove any DC offset
voltage from the waveform.
Press the Impedance softkey to set the input impedance of the oscilloscope channel to 1 MΩ to
make accurate measurements.
Press the Probe softkey to display the channel's probe menu.
Press the Units softkey to select Amps as the measurement unit for the connected probe.