Fuse Locations - Agilent Technologies G1289 Operating Manual

Headspace sampler
Table of Contents

Advertisement

Maintaining the Instrument

Fuse Locations

Fuse Locations
Refer to Figure 15 and Table 8 below for fuse locations and part numbers.
WARNING
Fuse replacement should be performed only by qualified service personnel.
P S 86.5 0 C S 20 06/A H S S
R 2
IC 2
R 8
IC 1
R 7
1
1
2
3
R 6
4
5
6
R 3
7
8
9
R 4
10
1 1
K 1
R 5
2 4V d c
1 4
K 2
6 8
2 4V
16 13 11 9
J2
D 2
D 1
Figure 15.
Fuse locations on the power supply
Table 8.
Fuse Location
F1
F1
F2
F3, F4
F5
F6
Power cord fuse holder
Power cord fuse holder
88
C 4
P D 2
+
~
+
Q 1
~
R 1
C 3
- -
D Z 2
F2
D S 2
F L 6
F L 2
D S 1
F6
D S 3
D S 4
1
4
7
10
2
5
8
11
12
0
3
6
9
12
J3
4.5A
1
4
7
10
2
5
8
11
3
12
6
9
J5
S
6
3
J1
5
2
4
1
Power supply board
Fuse Part Numbers
C 5
C 2
P D 3
+
C 1
~
+
+
+
~
C 8
C 6
C 7
- -
P D 1
F5
F3
F4
F L 5
F L 3
F L 4
19
0
15
0
15
0
15
1.2A
1.2A
300m A
T 1
0
110
0
110
J4
3
2
1
F1
F L 1
Voltage
Description
230
0.5 A, 5×20
115
1 A, 5×20 T
all
2 A, 5×20 T
all
0.5 A, 5×20
all
2 A, 5×20 T
all
6 A, 5×20 F
230
3 A, 5×20 T
115
6 A, 5×20 F
Power cord fuse
holder
Part No.
384-1695004-HSP
384-1595011-HSP
384-1695001-HSP
384-1695004-HSP
384-1695001-HSP
384-1295001-HSP
384-1695003-HSP
384-1295001-HSP