Vial Not Raised and Lowered in Oven

Probable Causes
Belt on the vial elevator is loose or broken.
Vial elevator motor is at fault. Check the voltage on J9 pins 3 and 9 on the
main board: V > 20 V or < –20 V.
Suggested Actions
1.
Check the belt. Replace it if necessary. See your service manual or contact
Agilent service.
2.
Exchange the motor.
3.
Call Agilent service.

Vial Not Loaded/Vial Retained Warning

Probable Causes
Robotic arm microswitch out of adjustment.
Vial remains in robotic arm.
Suggested Actions
Contact Agilent service.

Vial Jam Problems (G1289/G1290 only)

In certain situations, vials will jam when the unit attempts to drop one vial on
top of another in the oven or tries to place a vial in the sample tray when a vial
is already in that location. Vial jam problems can also result from misalignment
of the robotic arm or oven carousel. This may cause the motor for the oven
carousel or the robotic arm to jam.
Vial Jam Problems (G1289/G1290 only)
Troubleshooting
121