Zone Temps

Press Zone Temps to set temperature values for various heated zones in the
sampler. Pressing Zone Temps brings up the following top–level display:
G1289/G1290
G1883
Use the cursor keys to scroll to LOOP or TR. LINE (Transfer Line) to set
these temperature values. The allowed range of values is as follows:
Heated Zone
Allowed Setpoint Range*
G1883
Vial
40–200 in 1°C increments
Loop
50–200 in 1°C increments
Tr. Line
50–220 in 1°C increments
G1289/G1290
Oven
0–200 in 1°C increments**
Loop
0–220 in 1°C increments**
Tr. Line
0–220 in 1°C increments**
* The minimum oven temperature for all models is 40°C or 15°C above ambient,
whichever is greater.
** Do not enter a value below the minimum oven operating temperature, even though
temperatures between 0–39°C are accepted.
If a Zone Temp is changed during a run, the sampler will not display NOT
READY, but it will adjust the zone to the new temperature value.
As a general rule, the temperatures for the loop and transfer line should be
set higher than the oven temperature to avoid condensation. Set the loop
Actual
OVEN
100
Actual
VIAL
100
Keys and Their Functions
Setpoint
{120 }
Setpoint
{120 }
Default Value (°C)
70
50
50
70
80
90
The Keypad
31