Recovery From An Automation Error (G1289/G1290) - Agilent Technologies G1289 Operating Manual

Headspace sampler
Table of Contents

Advertisement

2.
To remove a vial from the robotic arm, see "Vial Not Loaded/Vial Retained
Warning" on page 121.
See also "Vial Jam Problems (G1289/G1290 only)" on page 121.
VIAL RETAINED
Probable Causes
The 7694 HS was reset with a vial still held by the robotic arm. This may
happen after a power failure or a manual reset.
Suggested Actions
1.
WARNING
Vials in the oven may be hot.
2.

Recovery from an Automation Error (G1289/G1290)

If an automation error occurs, there are two recovery modes.
If you want to reload the default method, hold Clear during power-up. The
following message is displayed.
If you want to reset the counter, hold Adv Fctn during power–up.
Power cycle the headspace sampler and hold down the clear key until the
tray is reset.
Manually remove the vial that is held in the robotic arm. See "Vial Not
Loaded/Vial Retained Warning" on page 121, or remove the cover from
the vertical loader and pull the vial out of the gripper.
Power cycle the headspace sampler and hold down the clear key until the
tray is reset.
LOAD DEFAULT METHOD
TRAY RESET
Recovery from an Automation Error (G1289/G1290)
Error Messages
133