Chromatography Problems; No Peaks - Agilent Technologies G1289 Operating Manual

Headspace sampler
Table of Contents

Advertisement

Troubleshooting
This chapter discusses the possible causes and the solutions for
chromatography problems or instrument failures.

Chromatography Problems

No Peaks

Probable Causes
Suggested Actions
1.
2.
116
Vial cracked
Vial cap leaking
Sample vial empty.
10 ml vial used without adapter.
Method not sensitive enough.
Leaks in the system.
Needle broken/clogged.
Sample condensed in the loop or transfer line.
Sample loop not filled with sample.
Inject time too short.
Sample did not create headspace.
Incorrect vial sampled.
Vial not punctured (may be due to mechanical failure or vial elevator jam).
Transfer line plugged.
Check that the proper vial is sampled and that the vial contains sample.
Use adapters with 10 ml vials.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

7694G1883G1290

Table of Contents