Advanced Function Descriptions - Agilent Technologies G1289 Operating Manual

Headspace sampler
Table of Contents

Advertisement

Using the Adv Fctn Key

Advanced Function Descriptions

Advanced Function Descriptions
CONSTANT HEATING TIME
G1883
This function indicates that the 7694 is always in the constant heating time
mode.
DEFAULT METHOD
G1289/G1290
Pressing Adv Fctn, 7, and Enter displays the following message:
Pressing Enter again loads a default method into the headspace sampler. The
display shows the following flashing message:
before returning to a top–level display. The following parameters are loaded in
the default method:
Oven Temp
Loop Temp
Tr. Line Temp
GC Cycle Time
Vial Eq.Time
Pressuriz.Time
Loop Fill Time
Loop Eq.Time
Inject Time
First Vial
Last Vial
Shake
44
Adv Fctn, 2, Enter
Adv Fctn, 7, Enter
LOAD DEFAULT METHOD
METHOD LOADED
70°C
80°C
90°C
25 min
15 min
0.2 min
0.2 min
0.05 min
1.0 min
1
10
Off