Press Esc key to quit this file and back to file selection main screen.
➢
Stored file
6.2.2 L-Leq mode file
When the file has been identified by its identification number:
➢ Press OK key to access to E1 main screen results.
LUpk : level of maximum peak of sound pressure
•
LXYmax : maximum value of sound pressure level weighted temporally met since the beginning of the measurement.
•
LXYmin : minimum value of sound pressure level weighted temporally met since the beginning of the measurement.
•
LXeq : equivalent continuous level of acoustic pressure frequency weighted on measurement time.
•
LXE : level of exposure to sound frequency weighted
•
D : measurement time expressed in DD / HH: MM: SS.
•
% : percentage of saturation of the input stage.
•
➢ With scrolling knob key, go to E2 screen which reminds measurement configuration.
Date of measurement start
•
Date of measurement end
•
Measurement time expressed in DD / HH: MM: SS.
•
Measurement mode : L-Leq
•
Frequency weighting for L and Leq : Weig.
•
Time weighting for L : Cst
•
Integration time for leq : S/S (Start/Stop mode)
•
Frequency weighting of the level of peak pressure for L : LpK
•
➢ Press Esc key to quit this file and back to file selection main screen.
Stored file
6.2.3 Leq-St mode file
When the file has been identified by its identification number:
Read the stored files
E1 screen
E1 screen
E2 screen
E2 screen
Stored file
Stored file