2. Confirm that you want to shut down your receiver.
a. If you want to
shut
b. If you want to cancel the shutdown, press the
the
Main
Menu.
To turn the receiver back on and resume communication with the transmitter, press the
may take up to 20 seconds for the display to turn back on.
Remember that your alerts and low glucose alarm will not work when the receiver is shut down.
Shutting down the receiver does not extend the sensor life beyond 7 days. Your sensor session will
stop 7 days after you started the sensor session.
6.11 THE DEXCOM G4 PLATINUM SYSTEM AND WATER
Your sensor is water resistant when showering, bathing or swimming if the transmitter is fully snapped in.
The sensor has been tested to be water resistant when submerged for up to 8 feet and up to 24 hours.
Keep the receiver dry. Do not spill fluids on it or drop it into fluids. Keep the micro USB port cover closed
to help prevent fluid from getting inside the receiver. Wireless communication does not work well
through water so the range is much less if you are in a pool, bathtub or water bed.
If your receiver gets wet, make sure the
the Try It option in the
Profiles
down, press the
LEFT
speaker
menu. See Chapter 9, Section 9.3, Alert Profiles.
button to highlight "OK," and then press the
button (with "Cancel" highlighted) to return to
SELECT
and vibrate mode are still working. You can do this using
37
button.
SELECT
button. It
SELECT
Need help?
Do you have a question about the G4 PLATINUM and is the answer not in the manual?
Questions and answers