Dexcom G4 PLATINUM User Manual page 103

Continuous glucose monitoring system
Hide thumbs Also See for G4 PLATINUM:
Table of Contents

Advertisement

Electrical Fast
Transient /burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Input
Lines
IEC 61000-4-11
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Receiver Product Specifications
Part Number
Reading Frequency
Dimensions
Weight
Receiver Input
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5% U
T
(> 95% dip in U
) for 0.5
T
cycle
40% U
T
(60% dip in U
) for 5
T
cycles
70% U
T
(30% dip in U
) for 25
T
cycles
< 5% U
T
(> 95% dip in U
) for 5
T
sec
3 A/m
MT22495
Every 5 minutes
Length: 4.0 inches
Width: 1.8 inches
Thickness:
0.5 inches
2.4 ounces
5V DC, 1A
Not applicable
Battery operated
Not applicable
Battery operated
Not applicable
Battery operated
3 A/m
103
relative humidity should
be at least 30%
Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.