Leak and Waste Handling
Figure 9
The waste tube connected to the leak pan outlet on each of the bottom
instruments guides the solvent to a suitable waste container.

Leak Sensor

Solvent incompatibility
C AU T I O N
The solvent DMF (dimethyl formamide) leads to corrosion of the leak sensor. The
material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
22
Infinity II Two Stack Leak Waste Concept (bench installation)
Do not use DMF.