Gas Connections

Table 2. Summary of the required gases for the Orbitrap Fusion Series MS
Type
Recommended purity
Helium
UHP (99.999%)—Has less than
1.0 ppm each of water, oxygen,
and total hydrocarbons.
Nitrogen
UHP (99.999%)—Has less than
3.0 ppm each of water, oxygen,
and total hydrocarbons.
HP (99%)
a
Approximate value when operating 24 hours and 7 days a week
Thermo Scientific
To connect the gas lines to the back of the Orbitrap Fusion Series MS, follow the next
procedure. Make sure that the ultra-high-purity (UHP) and high-purity (HP) supply gases are
delivered with the necessary pressure and purity as listed in
Note Connections and gas delivery systems might vary. You are responsible for supplying
any additional fittings or connections necessary during installation. If your system
includes additional devices that require their own gas connections, refer to those
instruments' manuals.
CAUTION You must use either copper (provided) or stainless steel for the UHP gas lines.
CAUTION After completing these connections, route the tubing so that they are not a trip
hazard.
Required
Typical daily
pressure
consumption
275 ±70 kPa
9.22 L
(40 ±10 psi)
(0.325 ft
345 ±70 kPa
46 L (1.62 ft
(50 ±10 psi)
690 ±140 kPa
11 500–26 700 L
(100 ±20 psi)
(406–943 ft
Table
2.
Function
a
Linear ion trap damping gas
3
)
and collision gas
3
)
Venting gas, ion routing
multipole (IRM) collision
gas, and make-up gas and
reagent carrier gas for the IC
and ETD configurations
API source auxiliary gas,
3
)
sheath gas, and sweep gas
Orbitrap Fusion Series Getting Connected Guide
2
5