5
Electrical Connections
To connect the MS and data system cables and power cords
Connect the Ethernet cables from
the Ethernet switch to the computer's
LC/MS Ethernet network card and the
MS.
1
(Not shown) Connect the contact
closure cable and the Ready Out
cable (not provided) between the
MS and the external device. See
Chapter 6, "External Devices."
2
3
Connect the USB cables from the
MS to the syringe pump and the
modular valve.
1
The back of the data system computer might not be as shown.
14
Orbitrap Fusion Series Getting Connected Guide
IMPORTANT Before turning on the MS, make sure to complete all of the system
connections described in this guide.
Tip For troubleshooting purposes, you might want to record which devices connect to
which ports on the Ethernet switch.
1
5
(Not shown) Connect both ends of the power cords
(for the computer, Ethernet switch, modular valve,
syringe pump, MS, and forepumps).
IMPORTANT Plug the power cords for the MS and
forepumps into separate, dedicated 230 Vac electrical
outlets—that is, circuits A, B, and C for a
dual-forepump system.
Note If the modular valve's LED remains off, press its
front Remote button to exit local mode.
4
Connect the relay control
cable between the MS
(V65 Pump Enable) and the
forepumps.
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading