Table 3. Pin-out descriptions for the communication connectors (Sheet 2 of 3)
Pin
Name
4
Start In
5
Ready Out
6
Injection Hold
8
RO/IH
Analog Input
The two analog channels connect to two separate 12-bit analog-to-digital converters (ADC)
for on-demand conversion of the input voltage. The conversion rate depends on the MS
rate.
1
Chassis
3, 4
2V Max:
+ (positive, pin 3) and
– (negative, pin 4)
5
Chassis
7, 8
10V Max:
+ (positive, pin 7) and
– (negative, pin 8)
Description
Receives the start signal from the contact closure
connection of a connected external device.
To activate this signal, the external device must pull the
signal either low (below 0.75 Vdc) or high (above
2.4 Vdc), depending on the polarity, for at least
100 ms by using a relay, an open-collector driver, or a
similar device that connects between pins 4 and 1.
Note In the Instrument Configuration window, set
the contact closure signal to "High-to-low edge" or
"Low-to-high edge," whichever matches the setting
for the connected external device.
Provides a relay-driven programmable output signal to
the connected external device. The relay opens when a
method starts and closes when the method finishes.
Output: Maximum 24 Vdc, 3 A
Provides a relay-driven programmable output signal to
the connected external device, such as a fraction
collector.
Output: Maximum 24 Vdc, 3 A
Common (return) connection for the Ready Out and
Injection Hold pins.
Earth ground (for pins 3 and 4).
(Channel 2) Provides a connection for an external
device, such as an LC instrument.
Input: 0–2 Vdc (voltage clamps at 5 Vdc)
Earth ground (for pins 7 and 8).
(Channel 1) Provides a connection for an external
device, such as an LC instrument.
Input: 0–10 Vdc (voltage clamps at 15 Vdc)