Section 11: Troubleshooting
Although service and repair of the Cavitron
should be performed by DENTSPLY
are some basic trouble shooting procedures that will help avoid
unnecessary service calls. Generally, check all lines and connections
to and from the System, a loose plug or connection will often
create problems. Check the settings on the System's knobs.

11.1 Troubleshooting Guide

System will not operate:
(Power Indicator Light is not lighted.)
1. Check that the Power Control knob is in the ON position,
and that the detachable Power Cord is fully seated in the
receptacle on back of the Cavitron
2. Check that the System's three-prong plug is fully seated in an
appropriate AC receptacle, and that AC current is present.
3. Check that the DC power plug is properly inserted into the
DC input located on the back of the System.
(Power Indicator Light is lighted.)
1. Check that the Foot Control Connector is fully seated in the
Foot Control Receptacle on the back of the System.
2. Check that the Handpiece Cable Assembly is fully seated in
the Handpiece Receptacle.
ENGLISH • 14
®
Select™ SPS™
personnel, the following
®
Select™ SPS™ System.
®
System operates:
(No lavage flow at insert tip.)
1. Assure that Lavage control is properly adjusted.
2. Check that irrigant supply control valve/s (dental office
water supply) are open.
3. Check that the Handpiece Cable Assembly is fully seated
in the Handpiece Receptacle.
4. Check to see if Water Line or Pump Lavage Filter is
discolored—if so, replace.
5. Check to assure that liquid is available in reservoir bottle.

11.2 Technical Support and Repairs

For technical support and assistance call 1-800-989-8826 Monday
through Friday, 8:00 AM to 5:00 PM (Eastern Time). For other
areas, contact your local DENTSPLY
representative.
®