none of the Tap-On
Foot Pedals operate the system,
TM
continue to the next step.
2.
Resynchronize one Tap-On
Foot Pedal to the
TM
system (see Section 7.10 Tap-On
Foot Pedal
TM
Synchronization).
Symptom:
System operates: No water flow to insert tip or
handpiece overheats
1.
Assure that handpiece lavage control is properly
adjusted.
2.
Check for clogged insert. Replace insert if necessary.
3.
Check that dental office water supply valves are open.
4.
If the system is connected to DualSelect Dispensing
System, check that fluid level in the selected bottle is
sufficient. Make sure valves are open when using
external water source.
5.
Check that the water line filter is clean. Replace
filter if needed.
Symptom:
System operates: No insert cavitation
1.
Check that the Power Level Control is not in Rinse
Mode.
2.
Check the insert for damage and that it is properly
installed in the handpiece.
3.
Check that the handpiece is properly installed to the
cable assembly.
4.
Verify that the soft nozzle grip is flush with the hard
plastic of the insert port.
5.
Turn the system’s Main Power Switch to the OFF (0)
position. Wait 5 seconds and turn the system back ON.
6.
If problem still exists, replace both “AA” batteries in
Tap-On
Foot Pedal with new “AA” batteries (Refer to
TM
Section 7.9) or connect the Auxiliary Tap-On
Foot
TM
Pedal Cable.
Symptom:
System operates: Purge Mode will not function –
icon flashing
1.
Check that there is no insert in the handpiece.
2.
Check that handpiece is properly installed to the cable
assembly.
Symptom:
System operates: Service Indicator blinking
• Fast Blinking (3 blinks per second)
– Indicates improper set-up
1. If insert is in the handpiece, remove. Verify the
handpiece is properly seated and hold the foot
control for 2 seconds. If blinking stops, the system
is ready for use. If blinking remains, continue to the
next step.
2. Attach a NEW handpiece and hold Tap-On
Foot
TM
Pedal for 2 seconds. If blinking stops, the system is
ready for use. Discard the old handpiece or return if
within warranty. If blinking remains, continue to the
next step.
3. Install and fully seat an insert into handpiece. Hold
Tap-On
Foot Pedal for 2 seconds. If blinking
TM
stops, the system is ready for use. If blinking
Page 21